- Measurement and conversion of electrical signals generated by thermocouples, RTDs and sensors with DC-current output.
- Up to 13 galvanically isolated measuring channels.
- Three-wire or four-wire RTD connection.
- Device conversion function – linear or square-root extracting.
- Individual setting each measuring channel.
- 2 NO relay outputs for signaling.
- Cold junction temperature compensation (internal or external sensor possible).
- Various modes for measurement results indication.
- Internal real-time clock with independent power supply.
- Interface RS-485, protocol Modbus (RTU).
- Internal archive capacity – 30720 records.
- Logging of measurement results and contingency situations with the possibility of subsequent transfer to a PC over RS-485 interface.
Technical specifications
Thermocouples | K, L, S, B |
RTDs:
|
Pt50, Pt100, Pt500 Cu50, Cu100 |
DC-current input signal range, mA | 0..5; 0..20; 4..20 |
Temperature measurement error (for thermocouples), °С:
|
±(0.25·10-2×|t|+0.5*) ±7.0 |
Temperature measurement error (for RTDs), °С | ±(0.2·10-2·(t+273)+0.1) |
Reduced DC-current signal measurement error, %:
|
±0.5 |
Least significant digit value | 0.001; 0.01; 0.1; 1 |
Supply voltage (depends on model) | 220VAC 50Hz 24VDC |
Operating temperature range | +5°С..+50°С |
Protection type (IP code) | IP41 |
Size (W×H×D), mm | 144×144×120 |
* – when internal sensor is used for cold junction temperature compensation it should be added 2.0 °С to error value
Ordering designation
Digital multi-channel device IP-6M-TK-X–YYY:
X – number of channels taken from a variety 5, 9, 13;
YYY – supply voltage variants:
«24V» – 24 VDC;
no marking – 220 VAC 50Hz.
Model IP-6M-TK-X-24V is equipped with a AC/DC converter 220VAC/24VDC.
Example. Digital multi-channel device IP-6M-TK-9-24V – device with 9 measuring channels, 24 VDC supply voltage.